DICOM Conformance Statement
for ExactVu™ 2.8
other versions:
v.2.0,
v.2.5,
v.2.5.3,
v.2.5.5,
v.2.8,
v.2.9
Revision 1.0
This is a DICOM Conformance Statement for ExactVu 2.8. ExactVu provides the capability to query modality worklist from a RIS scheduler, store ultrasound studies to, and query/retrieve MR studies including GSPS from a DICOM Image Manager. It can also read MR studies from a USB storage device or a CD/DVD through a USB port. The following DICOM SOP classes are supported by ExactVu:
Table Network Services
SOP Classes |
User of Service (SCU) |
Provider of Service (SCP) |
Transfer |
||
US Image Storage |
Yes |
No |
Ultrasound Multiframe Image Storage |
Yes |
No |
MR Image Storage |
No |
Yes |
Enhanced MR Image Storage |
No |
Yes |
MR Spectroscopy Storage |
No |
Yes |
Grayscale Softcopy Presentation State Storage | No |
Yes |
Secondary Capture Image Storage | No |
Yes |
Key Object Selection Document | No |
Yes |
X-Ray Radiation Dose SR | No |
Yes |
Chest CAD SR | No |
Yes |
Mammography CAD SR | No |
Yes |
Basic Text SR | No |
Yes |
Enhanced SR | No |
Yes |
Comprehensive SR | No |
Yes |
Raw Data Storage | No |
Yes |
Procedure Log Storage | No |
Yes |
Multiframe Grayscale Byte Secondary Capture Image Storage | No |
Yes |
Multiframe Grayscale Word Secondary Capture Image Storage | No |
Yes |
Multiframe Single Bit Secondary Capture Image Storage | No |
Yes |
Multiframe True Color Secondary Capture Image Storage | No |
Yes |
Spatial Fiducials Storage | No |
Yes |
Spatial Registration Storage | No |
Yes |
Standalone Curve Storage | No |
Yes |
Hardcopy Color Image Storage | No |
Yes |
Hardcopy Grayscale Image Storage | No |
Yes |
Standalone Modality LUT Storage | No |
Yes |
Standalone Overlay Storage | No |
Yes |
Blending Softcopy Presentation State Storage | No |
Yes |
Color Softcopy Presentation State Storage | No |
Yes |
Encapsulated PDF Storage | No |
Yes |
Pseudocolor Softcopy Presentation Stage Storage | No |
Yes |
Deformable Spatial Registration Storage | No |
Yes |
Segmentation Storage | No |
Yes |
Workflow Management |
||
Modality Worklist Information Model - FIND |
Yes |
No |
Query/Retrieve |
||
Study Root Query/Retrieve Information Model – FIND |
Yes |
No |
Study Root Query/Retrieve Information Model – MOVE |
Yes |
No |
Table Media Services
Media Storage Application Profile |
Read/Write Files |
Compact Disk – Recordable |
|
General Purpose CD-R Interchange |
FSR |
General Purpose CD-R Interchange with JPEG |
FSR |
General Purpose CD-R Interchange with J2K |
FSR |
CT/MR Studies on CD-R |
FSR |
DVD |
|
General Purpose DVD Interchange |
FSR |
General Purpose DVD Interchange with JPEG |
FSR |
General Purpose DVD Interchange with J2K |
FSR |
CT/MR Studies on DVD Media |
FSR |
USB and Flash Memory |
|
General Purpose USB |
FSR |
General Purpose USB with JPEG |
FSR |
General Purpose USB with J2K |
FSR |
Network Drive |
|
A network drive mapped by a network-shared folder |
FSR |
Note:
Note: ExactVu serves as a DICOM File-Set Reader (FSR); however, it only reads MR studies with some extra specifications for its FusionVu functionality. Network Drive is not a standard Media storage Application Profile. ExactVu handles the network drive in the same way as a CD-R or DVD and serves as a DICOM File-Set Reader (FSR).
1 Conformance Statement Overview
2 Table of Contents
3 Introduction
3.1 Audience
3.2 Scope and field of Application
3.3 Definition, Terms and Abbreviation
4 Networking
4.1 Implementation Model
4.2 AE Specifications
4.3 Network Interfaces
4.4 Configuration
5 Media Interchange
5.1 Implementation Model
5.2 AE Specifications
6 Transformation of DICOM to CDA
7 Support of Character Sets
8 Security
9 Annexes
This DICOM Conformance Statement is intended for following audiences:
It is the intent of this document to provide an unambiguous specification for DICOM implementations of ExactVu. This specification includes a DICOM Conformance Statement and is necessary to ensure proper processing and interpretation of ExactVu medical data exchanged using DICOM.
The use of the DICOM Conformance Statement, in conjunction with the DICOM Standards, is intended to facilitate communication with the ExactVu high resolution micro-ultrasound system. However, by itself, it is not sufficient to ensure that inter-operation will be successful.
The reader of this DICOM Conformance Statement should be aware of the following important issues:
Definitions, terms and abbreviations used in this document are defined within the different parts of the DICOM standard. Abbreviations and terms used are as follows:
Term |
Definition |
EVU |
High-resolution micro-ultrasound device from Exact Imaging Inc. |
AE |
DICOM Application Entity |
ASCE |
Association Control Service Element |
FSR |
File-Set Reader |
GSPS or PR |
Grayscale Softcopy Presentation State |
IOD |
Information Object Definition |
ISO |
International Standard Organization |
MR or MRI |
Magnetic Resonance Imaging |
PDU |
Protocol Data Unit |
SCU |
Service Class User |
SCP |
Service Class Provider |
SOP |
Service-Object Pair |
TCP/IP |
Transmission Control Protocol/Internet Protocol |
UID |
Unique Identifier |
ExactVu implements DICOM verification SCU, Storage SCU, Storage Commitment SCU. It also implements Query /Retrieve - MOVE SCU and a temporary Storage SCP for image retrieving. Below is the application data flow diagram.
Figure 1: DICOM Standard Interface
All communications and image transfer with remote application is accomplished using the
DICOM protocol over a network using the TCP/IP protocol stack.
The DICOM verification SCU (i.e., connectivity test) is available for testing and validation purposes of remote AEs. ExactVu opens an association and sends C-ECHO request to verify specified DICOM SCP node. Upon receiving the response from SCP, or in case of failure, it closes the connection.
DICOM Storage SCU is available to send local Ultrasound studies, which are selected either by the user or background archiving to the configured remote DICOM Storage SCP. ExactVu opens an association and send a single-framed or multi-framed image to the specific SCP. Upon receiving the response or in case of failure, it closes the connection.
ExactVu searches different level information of the DICOM studies as a study-root Query/Retrieve - FIND SCU from the remote AE, i.e., a DICOM query/retrieve SCP by sending a C-FIND request and receiving the response.
ExactVu retrieves the DICOM SOP instances as a study-root Query/Retrieve - MOVE SCU from the remote AE, i.e., a DICOM query/retrieve SCP by sending a C-MOVE request and receiving the response. Once the response is successful, it will setup a temporary Storage SCP as described in 4.1.2.6.
ExactVu initiates a temporary Storage SCP to accept the SOP instance sending, which is initiated by a C-MOVE request, from the remote AE. Once the receiving task is done or times out, the Storage SCP will be destroyed.
ExactVu searches modality worklist as an MWL FIND SCU from the remote AE, i.e., a DICOM Modality Worklist Management SCP by sending an MWL FIND request and receiving the response.
Not applicable.
Verification provides e Standard Conformance to the following SOP Class:
SOP Class Name |
SOP Class UID |
SCU |
SCP |
Verification |
1.2.840.10008.1.1 |
Yes |
No |
ExactVu acts as verification SCU AE to test the DICOM connectivity to the remote AE.
The DICOM standard application context name for DICOM 3.0 is always proposed.
Application Context Name |
1.2.840.10008.3.1.1.1 |
This version of implementation only supports one simultaneous association.
Maximum number of simultaneous Associations |
1 |
This version of implementation does not support asynchronous communication (multiple outstanding transactions over a single Association).
Maximum number of outstanding asynchronous transactions |
1 |
ExactVu uses the following implementation identifying parameters:
Implementation Class UID |
2.16.124.113639.1.0.2.5.2.0 |
Implementation Version |
MergeCOM3_5_6_0 |
Note the Implementation Version may be updated with the release of the product software.
ExactVu initiates Associations only for the purpose of verifying a DICOM connection.
Presentation Context Table |
|||||
Abstract Syntax |
Transfer Syntax |
Role |
Extended Negotiation |
||
Name |
UID |
Name |
UID |
||
Verification |
1.2.840.10008.1.1 |
Implicit VR Little Endian |
1.2.840.10008.1.2 |
SCU |
None |
Explicit VR Little Endian |
1.2.840.10008.1.2.1 |
ExactVu provides standard conformance to the DICOM Verification Service Class as an SCU.
The status code for the C-ECHO is as following:
Code |
Status |
Meaning |
0000 |
Success |
The C-ECHO request is accepted. |
Not applicable.
Storage AEs provide Standard Conformance to the following SOP Classes:
SOP Class Name |
SOP Class UID |
SCU |
SCP |
US Image Storage |
1.2.840.10008.5.1.4.1.1.6.1 |
Yes |
No |
Ultrasound Multiframe Image Storage |
1.2.840.10008.5.1.4.1.1.3.1 |
Yes |
No |
The DICOM standard application context name for DICOM 3.0 is always proposed.
Application Context Name |
1.2.840.10008.3.1.1.1 |
The Storage SCU of the Storage AE initiates one association at a time for the destination.
Maximum number of simultaneous Associations |
1 |
The Storage Commitment SCU opens one association at a time to receive N-EVENT-REPORT notification.
Maximum number of simultaneous Associations |
1 |
This version of implementation does not support asynchronous communication (multiple outstanding transactions over a single Association).
Maximum number of outstanding asynchronous transactions |
1 |
ExactVu uses the following implementation identifying parameters:
Implementation Class UID |
2.16.124.113639.1.0.2.5.2.0 |
Implementation Version |
MergeCOM3_5_6_0 |
Note the Implementation Version may be updated with the release of the product software.
The images of a study selected manually by the user or automatically by the background archiving system can be sent to a remote AE. If Storage SCU successfully establishes an Association to a remote AE, it will transfer each SOP instance one after another via the open Association. After successful or a failed sending, the association will be closed. If the sending is successful and the Storage Commitment feature is turned on, Storage Commitment SCU will transmit setup another association and send a Storage Commitment request (N-ACTION), waiting for a N-EVENT-REPORT at the specific TCP port until succeeded or time out, then close the association.
Presentation Context Table |
|||||
Abstract Syntax |
Transfer Syntax |
Role |
Extended Negotiation |
||
Name |
UID |
Name |
UID |
||
US Image Storage |
1.2.840.10008.5.1.4.1.1.6.1 |
JPEG 2000 Image Compression (Lossless Only) |
1.2.840.10008.1.2.4.90 |
SCU |
None |
US Multiframe Image Storage |
1.2.840.10008.5.1.4.1.1.3.1 |
JPEG 2000 Image Compression (Lossless Only) |
1.2.840.10008.1.2.4.90 |
SCU |
None |
ExactVu provides standard conformance to the DICOM Standard Storage Service Class as SCU for the storage SOP classes of US Image Storage and US Multiframe Image Storage as listed in the table in section 4.2.2.3.1.2. ExactVu sends the following attributes in C_STORE_RQ. All the mandatory attributes are sent.
Group |
Element |
Description |
0008 |
0005 |
Specific Character Set |
0008 |
0008 |
Image Type |
0008 |
0016 |
SOP Class UID |
0008 |
0018 |
SOP Instance UID |
0008 |
0020 |
Study Date |
0008 |
0021 |
Series Date |
0008 |
0022 |
Acquisition Date |
0008 |
0023 |
Content Date |
0008 |
0030 |
Study Time |
0008 |
0031 |
Series Time |
0008 |
0032 |
Acquisition Time |
0008 |
0033 |
Content Time |
0008 |
0050 |
Accession Number |
0008 |
0060 |
Modality |
0008 |
0070 |
Manufacturer |
0008 |
0080 |
Institution Name |
0008 |
0090 |
Name of the patient's referring physician |
0008 |
0201 |
Time zone offset from UTC |
0008 |
1030 |
Study Description |
0008 |
103E |
Series Description |
0008 |
1090 |
Manufacturer's Model Name |
0010 |
0010 |
Patient's Name |
0010 |
0020 |
Patient ID |
0010 |
0021 |
Issuer of Patient ID |
0010 |
0030 |
Patient's Birth Date |
0010 |
0040 |
Patient's Sex |
0020 |
000D |
Study Instance UID |
0020 |
000E |
Series Instance UID |
0020 |
0010 |
Study ID |
0020 |
0011 |
Series Number |
0020 |
0013 |
Instance Number |
0020 |
9221 |
Dimension Organization Sequence |
>0020 |
9164 |
Dimension Organization UID |
0020 |
9222 |
Dimension Index Sequence |
>0020 |
9164 |
Dimension Organization UID |
>0020 |
9165 |
Dimension Index Pointer |
>0020 |
9167 |
Functional Group Pointer |
>0020 |
9421 |
Dimension Description Label |
0028 |
0002 |
Samples per Pixel |
0028 |
0004 |
Photometric Interpretation |
0028 |
0006 |
Planar Configuration |
0028 |
0010 |
Rows |
0028 |
0011 |
Columns |
0028 |
0030 |
Pixel Spacing |
0028 |
0100 |
Bits Allocated |
0028 |
0101 |
Bits Stored |
0028 |
0102 |
High Bit |
0028 |
0103 |
Pixel Representation |
0028 |
6010 |
Representative Frame Number |
0028 |
6020 |
Frame Numbers of Interest |
0028 |
6022 |
Frame of Interest Description |
0028 |
6023 |
Frame of Interest Type |
5200 |
9229 |
Shared Functional Groups Sequence |
>0028 |
9110 |
Pixel Measures Sequence |
>>0028 |
0030 |
Pixel Spacing |
5200 |
9230 |
Per Frame Functional Groups Sequence |
>0020 |
9111 |
Frame Content Sequence |
>>0018 |
9074 |
Frame Acquisition Date Time |
>>0018 |
9151 |
Frame Reference Date Time |
>>0020 |
9056 |
Stack ID |
>>0020 |
9057 |
In Stack Position Number |
>>0020 |
9128 |
Temporal Position Index |
>>0020 |
9157 |
Dimension Index Values |
7FE0 |
0010 |
Pixel Data |
Not applicable.
SOP Class Name |
SOP Class UID |
SCU |
SCP |
Study Root Query/Retrieve Information Model – FIND |
1.2.840.10008.5.1.4.1.2.2.1 |
Yes |
No |
The DICOM standard application context name for DICOM 3.0 is always proposed.
Application Context Name |
1.2.840.10008.3.1.1.1 |
This version of implementation only supports one simultaneous association.
Maximum number of simultaneous Associations |
1 |
This version of implementation does not support asynchronous communication (multiple outstanding transactions over a single Association).
Maximum number of outstanding asynchronous transactions |
1 |
ExactVu uses the following implementation identifying parameters:
Implementation Class UID |
2.16.124.113639.1.0.2.5.2.0 |
Implementation Version |
MergeCOM3_5_6_0 |
Note the Implementation Version may be updated with the release of the product software.
Query/Retrieve - FIND SCU initiates a new association when the user performs the query action from the user interface. Once the query succeeds or fails, the association will be closed.
Presentation Context Table |
|||||
Abstract Syntax |
Transfer Syntax |
Role |
Extended Negotiation |
||
Name |
UID |
Name |
UID |
||
Study Root Query/Retrieve Information Model - FIND |
1.2.840.10008.5.1.4.1.2.2.1 |
Implicit VR Little Endian |
1.2.840.10008.1.2 |
SCU |
None |
Explicit VR Little Endian |
1.2.840.10008.1.2.1 |
ExactVu provides standard conformance to the DICOM Study Root Query/Retrieve Information Model – FIND SOP SCU Class.
The request identifiers specified by ExactVu are listed in the following table:
Name |
Tag |
Type of Matching |
Patient’s Name |
0010,0010 |
S, *, U |
Patient ID |
0010,0020 |
S, *, U |
Patient's Birth Date |
0010,0030 |
S, *, U, R |
Study Date |
0008,0020 |
S, *, U, R |
Accession Number |
0008,0050 |
S, *, U |
Types of Matching:
The types of Matching supported by the C-FIND SCU. An "S" indicates the identifier attribute uses Single Value Matching, an "R" indicates Range Matching, a n"*"indicates wild card matching, a 'U' indicates Universal Matching, and an 'L' indicates that UID lists are sent. "NONE" indicates that no matching is supported, but that values for this Element are requested to be returned (i.e., universal matching), and "UNIQUE" indicates that this is the Unique Key for that query level, in which case Universal Matching or Single Value Matching is used depending on the query level
Note: ExactVu only query series level with request modality identifier as MR or PR for FusionVu feature.
Query/Retrieve - FIND SCU does not accept associations.
SOP Class Name |
SOP Class UID |
SCU |
SCP |
Study Root Query/Retrieve Information Model – MOVE |
1.2.840.10008.5.1.4.1.2.2.2 |
Yes |
No |
The DICOM standard application context name for DICOM 3.0 is always proposed.
Application Context Name |
1.2.840.10008.3.1.1.1 |
This version of implementation only supports one simultaneous association.
Maximum number of simultaneous Associations |
1 |
This version of implementation does not support asynchronous communication (multiple outstanding transactions over a single Association).
Maximum number of outstanding asynchronous transactions |
1 |
ExactVu uses the following implementation identifying parameters:
Implementation Class UID |
2.16.124.113639.1.0.2.5.2.0 |
Implementation Version |
MergeCOM3_5_6_0 |
Note the Implementation Version may be updated with the release of the product software.
Query/Retrieve - MOVE SCU initiates a new association when the user performs MR study loading from the user interface. Once the C-MOVE responding succeeds, a temporary Storage SCP will be setup. the association will be closed after the retrieval success through the temporary Storage SCP.
Presentation Context Table |
|||||
Abstract Syntax |
Transfer Syntax |
Role |
Extended Negotiation |
||
Name |
UID |
Name |
UID |
||
Study Root Query/Retrieve Information Model - MOVE |
1.2.840.10008.5.1.4.1.2.2.2 |
Implicit VR Little Endian |
1.2.840.10008.1.2 |
SCU |
None |
Explicit VR Little Endian |
1.2.840.10008.1.2.1 |
ExactVu provides standard conformance to the DICOM Study Root Query/Retrieve Information Model – MOVE SOP SCU Class.
The retrieval is performed from the AE that was specified in the Retrieve AE attribute returned from the query performed by the FIND SCU. The instances are retrieved to local by specifying the destination as the AE Title of the temporary setup Storage SCP, which is described in section 4.2.5. This implies that the remote C-MOVE SCP must be preconfigured to determine the presentation address corresponding to the Storage SCP AE setup by ExactVu. The Storage SCP AE will accept storage requests addressed to it.
A retrieval can be performed at the STUDY, SERIES or IMAGE level depending on what level of entity has been selected. For FusionVu feature, ExactVu will perform a retrieval for instances in all PR(GSPS) series, then retrieve the GSPS referenced MR series.
The request identifiers for Query/Retrieve – Move SCU are listed in the table below:
Name |
Tag |
Unique, Matching or Return Key |
Study Level |
||
Study Instance UID |
0020,000D |
U |
Query/Retrieve - MOVE SCU does not accept associations.
SOP Class Name |
SOP Class UID |
MR Image Storage |
1.2.840.10008.5.1.4.1.1.4 |
Enhanced MR Image Storage |
1.2.840.10008.5.1.4.1.1.4.1 |
MR Spectroscopy Storage |
1.2.840.10008.5.1.4.1.1.4.2 |
Grayscale Softcopy Presentation State Storage |
1.2.840.10008.5.1.4.1.1.11.1 |
Secondary Capture Image Storage | 1.2.840.10008.5.1.4.1.1.7 |
Key Object Selection Document | 1.2.840.10008.5.1.4.1.1.88.59 |
X-Ray Radiation Dose SR | 1.2.840.10008.5.1.4.1.1.88.67 |
Chest CAD SR | 1.2.840.10008.5.1.4.1.1.88.65 |
Mammography CAD SR | 1.2.840.10008.5.1.4.1.1.88.50 |
Basic Text SR | 1.2.840.10008.5.1.4.1.1.88.11 |
Enhanced SR | 1.2.840.10008.5.1.4.1.1.88.22 |
Comprehensive SR | 1.2.840.10008.5.1.4.1.1.88.33 |
Raw Data Storage | 1.2.840.10008.5.1.4.1.1.66 |
Procedure Log Storage | 1.2.840.10008.5.1.4.1.1.88.40 |
Multiframe Grayscale Byte Secondary Capture Image Storage | 1.2.840.10008.5.1.4.1.1.7.1 |
Multiframe Grayscale Word Secondary Capture Image Storage | 1.2.840.10008.5.1.4.1.1.7.3 |
Multiframe Single Bit Secondary Capture Image Storage | 1.2.840.10008.5.1.4.1.1.7.1 |
Multiframe True Color Secondary Capture Image Storage | 1.2.840.10008.5.1.4.1.1.7.4 |
Spatial Fiducials Storage | 1.2.840.10008.5.1.4.1.1.66.2 |
Spatial Registration Storage | 1.2.840.10008.5.1.4.1.1.66.1 |
Standalone Curve Storage | 1.2.840.10008.5.1.4.1.1.9 |
Hardcopy Color Image Storage | 1.2.840.10008.5.1.1.30 |
Hardcopy Grayscale Image Storage | 1.2.840.10008.5.1.1.29 |
Standalone Modality LUT Storage | 1.2.840.10008.5.1.4.1.1.10 |
Standalone Overlay Storage | 1.2.840.10008.5.1.4.1.1.8 |
Blending Softcopy Presentation State Storage | 1.2.840.10008.5.1.4.1.1.11.4 |
Color Softcopy Presentation State Storage | 1.2.840.10008.5.1.4.1.1.11.2 |
Encapsulated PDF Storage | 1.2.840.10008.5.1.4.1.1.104.1 |
Pseudocolor Softcopy Presentation Stage Storage | 1.2.840.10008.5.1.4.1.1.11.3 |
Deformable Spatial Registration Storage | 1.2.840.10008.5.1.4.1.1.66.3 |
Segmentation Storage | 1.2.840.10008.5.1.4.1.1.66.4 |
The Storage SCP accepts but never initiates associations.
The DICOM standard application context name for DICOM 3.0 is always proposed.
Application Context Name |
1.2.840.10008.3.1.1.1 |
This version of implementation only supports one simultaneous association.
The Storage SCP AE listens to a specific port and accepts the associations initiated by the C-Store SCU on the PACS server side. For each Association a thread is created to receive SOP instances sent from remote AE, which is a sub-operation of ExactVu’s Query/Retrieve - MOVE. Although the simultaneous associations requested by the peer AE is not limited, the SCP doesn’t expect a lot of associations from a specific Storage SCU, i.e., the PACS server which provides SOP instances for sub-operations of a C-MOVE.
Maximum number of simultaneous Associations requested by the peer AE |
Not limited |
This version of implementation does not support asynchronous communication (multiple outstanding transactions over a single Association).
Maximum number of outstanding asynchronous transactions |
1 |
ExactVu uses the following implementation identifying parameters:
Implementation Class UID |
2.16.124.113639.1.0.2.5.2.0 |
Implementation Version |
MergeCOM3_5_6_0 |
Note the Implementation Version may be updated with the release of the product software.
Storage SCP does not initiate associations.
When the Storage SCP accepts an association, it will respond to the storage requests. If the Called AE Title does not match the pre- configured AE Title for the SCP of the application, the association will be rejected.
4.2.5.4.1 Activity - Receive Storage Request
4.2.5.4.1.1 Description and Sequencing of Activities
As instances are received, they are saved to the local file system and the records are saved by ExactVu application so that the SOP instances can be load for Fusion.
Presentation Context Table |
||||
Abstract Syntax |
Transfer Syntax |
|||
Name |
UID |
Name |
UID |
|
MR Image Storage |
1.2.840.10008.5.1.4.1.1.4 |
Implicit VR Little Endian |
1.2.840.10008.1.2 |
|
Explicit VR Little Endian |
1.2.840.10008.1.2.1 |
|||
Explicit VR Big Endian | 1.2.840.10008.1.2.2 | |||
Deflated Explicit VR Little Endian | 1.2.840.10008.1.2.1.99 | |||
JPEG Baseline (Process 1): Default Transfer Syntax for Lossy JPEG 8-bit Image Compression |
1.2.840.10008.1.2.4.50 | |||
JPEG Baseline (Processes 2 & 4): Default Transfer Syntax for Lossy JPEG 12-bit Image Compression (Process 4 only) |
1.2.840.10008.1.2.4.51 | |||
JPEG Lossless, Nonhierarchical (Processes 14) | 1.2.840.10008.1.2.4.57 | |||
JPEG Lossless, Nonhierarchical, First- Order Prediction |
1.2.840.10008.1.2.4.70 | |||
JPEG-LS Lossless Image Compression | 1.2.840.10008.1.2.4.80 | |||
JPEG-LS Lossy (Near- Lossless) Image Compression | 1.2.840.10008.1.2.4.81 | |||
JPEG 2000 Image Compression (Lossless Only) | 1.2.840.10008.1.2.4.90 | |||
JPEG 2000 Image Compression | 1.2.840.10008.1.2.4.91 | |||
Enhanced MR Image |
1.2.840.10008.5.1.4.1.1.4.1 |
Implicit VR Little Endian |
1.2.840.10008.1.2 |
|
Explicit VR Little Endian |
1.2.840.10008.1.2.1 |
|||
Explicit VR Big Endian | 1.2.840.10008.1.2.2 | |||
Deflated Explicit VR Little Endian | 1.2.840.10008.1.2.1.99 | |||
JPEG Baseline (Process 1): Default Transfer Syntax for Lossy JPEG 8-bit Image Compression |
1.2.840.10008.1.2.4.50 | |||
JPEG Baseline (Processes 2 & 4): Default Transfer Syntax for Lossy JPEG 12-bit Image Compression (Process 4 only) |
1.2.840.10008.1.2.4.51 | |||
JPEG Lossless, Nonhierarchical (Processes 14) | 1.2.840.10008.1.2.4.57 | |||
JPEG Lossless, Nonhierarchical, First- Order Prediction |
1.2.840.10008.1.2.4.70 | |||
JPEG-LS Lossless Image Compression | 1.2.840.10008.1.2.4.80 | |||
JPEG-LS Lossy (Near- Lossless) Image Compression | 1.2.840.10008.1.2.4.81 | |||
JPEG 2000 Image Compression (Lossless Only) | 1.2.840.10008.1.2.4.90 | |||
JPEG 2000 Image Compression | 1.2.840.10008.1.2.4.91 | |||
Secondary Capture Image |
1.2.840.10008.5.1.4.1.1.7 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Explicit VR Big Endian | 1.2.840.10008.1.2.2 | |||
Deflated Explicit VR Little Endian | 1.2.840.10008.1.2.1.99 | |||
JPEG Baseline (Process 1): Default Transfer Syntax for Lossy JPEG 8-bit Image Compression |
1.2.840.10008.1.2.4.50 | |||
JPEG Baseline (Processes 2 & 4): Default Transfer Syntax for Lossy JPEG 12-bit Image Compression (Process 4 only) |
1.2.840.10008.1.2.4.51 | |||
JPEG Lossless, Nonhierarchical (Processes 14) | 1.2.840.10008.1.2.4.57 | |||
JPEG Lossless, Nonhierarchical, First- Order Prediction |
1.2.840.10008.1.2.4.70 | |||
JPEG-LS Lossless Image Compression | 1.2.840.10008.1.2.4.80 | |||
JPEG-LS Lossy (Near- Lossless) Image Compression | 1.2.840.10008.1.2.4.81 | |||
JPEG 2000 Image Compression (Lossless Only) | 1.2.840.10008.1.2.4.90 | |||
JPEG 2000 Image Compression | 1.2.840.10008.1.2.4.91 | |||
MR Spectroscopy |
1.2.840.10008.5.1.4.1.1.4.2 |
Implicit VR Little Endian |
1.2.840.10008.1.2 |
|
Explicit VR Little Endian |
1.2.840.10008.1.2.1 |
|||
Raw Data | 1.2.840.10008.5.1.4.1.1.66 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Procedure Log | 1.2.840.10008.5.1.4.1.1.88.40 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Multi-frame Grayscale Byte Secondary Capture Image | 1.2.840.10008.5.1.4.1.1.7.2 | Implicit VR Little Endian |
|
|
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Multi-frame Grayscale Word Secondary Capture Image | 1.2.840.10008.5.1.4.1.1.7.3 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Multi-frame Grayscale Bit Secondary Capture Image | 1.2.840.10008.5.1.4.1.1.7.1 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Multi-frame True Color Secondary Capture Image | 1.2.840.10008.5.1.4.1.1.7.4 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Spatial Registration | 1.2.840.10008.5.1.4.1.1.66.1 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Spatial Fiducials | 1.2.840.10008.5.1.4.1.1.66.2 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Standalone Curve | 1.2.840.10008.5.1.4.1.1.9 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Hardcopy Color Image | 1.2.840.10008.5.1.1.30 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Hardcopy Grayscale Image | 1.2.840.10008.5.1.1.29 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Standalone Modality LUT | 1.2.840.10008.5.1.4.1.1.10 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Standalone Overlay | 1.2.840.10008.5.1.4.1.1.8 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Blending Softcopy Presentation State | 1.2.840.10008.5.1.4.1.1.11.4 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Color Softcopy Presentation State | 1.2.840.10008.5.1.4.1.1.11.2 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Encapsulated PDF | 1.2.840.10008.5.1.4.1.1.104. | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Pseudocolor Softcopy Presentation Stage | 1.2.840.10008.5.1.4.1.1.11.3 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Deformable Spatial Registration | 1.2.840.10008.5.1.4.1.1.66.3 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Segmentation | 1.2.840.10008.5.1.4.1.1.66.4 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Grayscale Softcopy Presentation State | 1.2.840.10008.5.1.4.1.1.11.1 | Implicit VR Little Endian | 1.2.840.10008.1.2 | |
Explicit VR Little Endian | 1.2.840.10008.1.2.1 | |||
Grayscale Softcopy Presentation State Storage |
1.2.840.10008.5.1.4.1.1.11.1 |
Implicit VR Little Endian |
1.2.840.10008.1.2 |
|
Explicit VR Little Endian |
1.2.840.10008.1.2.1 |
Storage SCP provides standard conformance to the Storage Service Class.
SOP Class Name |
SOP Class UID |
SCU |
SCP |
Modality Worklist Information Model - FIND |
1.2.840.10008.5.1.4.31 |
Yes |
No |
The DICOM standard application context name for DICOM 3.0 is always proposed.
Application Context Name |
1.2.840.10008.3.1.1.1 |
This version of implementation only supports one simultaneous association.
Maximum number of simultaneous Associations |
1 |
This version of implementation does not support asynchronous communication (multiple outstanding transactions over a single Association).
Maximum number of outstanding asynchronous transactions |
1 |
ExactVu uses the following implementation identifying parameters:
Implementation Class UID |
2.16.124.113639.1.0.2.5.2.0 |
Implementation Version | MergeCOM3_5_6_0 |
Note the Implementation Version may be updated with the release of the product software.
MWL - FIND SCU initiates a new association when the user performs the query action from the user interface. Once the query succeeds or fails, the association will be closed.
Presentation Context Table |
|||||
Abstract Syntax |
Transfer Syntax |
Role |
Extended Negotiation |
||
Name |
UID |
Name |
UID |
||
Modality Worklist Information Model - FIND |
1.2.840.10008.5.1.4.31 |
Implicit VR Little Endian |
1.2.840.10008.1.2 |
SCU |
None |
ExactVu provides standard conformance to the DICOM Modality Worklist Information Model – FIND SOP SCU Class.
The request identifiers specified by ExactVu are listed in the following table:
Name |
Tag |
Type of Matching |
Patient’s Name |
0010,0010 |
S, *, U |
Patient ID |
0010,0020 |
S, *, U |
Patient's Birth Date |
0010,0030 |
NONE |
Patient Sex |
0010,0040 |
S, *, U |
Study Instance UID | 0020,000D | NONE |
Scheduled Station AE Title | 0040,0001 | S |
Scheduled Procedure Step Start Date | 0040,0002 | S, R |
Scheduled Procedure Step Start Time | 0040,0003 | NONE |
Scheduled Performing Physician's Name | 0040,0006 | NONE |
Scheduled Procedure Step Description | 0040,0007 | NONE |
Scheduled Station Name | 0040,0010 | S, *, U |
Requested Procedure ID | 0040,1001 | NONE |
Specific Character Set | 0008,0005 | NONE |
Accession Number | 0008,0050 | NONE |
Modality |
0008,0060 |
S, *, U |
Types of Matching:
The types of Matching supported by the C-FIND SCU. An "S" indicates the identifier attribute uses Single Value Matching, an "R" indicates Range Matching, a n"*"indicates wild card matching, a 'U' indicates Universal Matching, and an 'L' indicates that UID lists are sent. "NONE" indicates that no matching is supported, but that values for this Element are requested to be returned (i.e., universal matching), and "UNIQUE" indicates that this is the Unique Key for that query level, in which case Universal Matching or Single Value Matching is used depending on the query level
Modality Worklist - FIND SCU does not accept associations.
ExactVu uses the MergeCOM-3 Advanced DICOM Toolkit to communicate over the TCP/IP protocol stack on any physical interconnection media supporting the TCP/IP stack. The Toolkit inherits the TCP/IP stack from the host operating system upon which it executes, i.e., Windows 7 for ExactVu.
Not applicable.
Any implementation's DICOM conformance may be dependent upon configuration, which takes place at the time of installation.
AE Titles on ExactVu can be configured from Preferences/Network Settings page. The parameters and default values are specified as following:
Service Class Name |
Role |
Remote Port |
Remote AE |
Local AE |
Local Port |
Storage |
SCU |
104 |
|
|
Not applicable |
Storage |
SCP |
104 |
|
|
105 |
Study Root Query/Retrieve |
SCU |
104 |
|
|
Not applicable |
Basic Worklist Management | SCU | 104 | Not applicable |
Note:
ExactVu implements DICOM File-Set Reader.
Figure 2: DICOM File-Set Reader Flow Diagram
ExactVu imports images and Presentation States from a removable media storage. The real-world activity “Import DICOM Objects from Media” is performed when the user requests to read a MR study list and/or images from a CD/DVD or USB storage device.
When the user selects to display MRI study list from a USB based CD/DVD drive, or from a USB storage device, ExactVu acts as a FSR to read the DICOMDIR from the DICOM file set on the media, i.e., CD/DVD or USB storage device, and build a study list to display to the user. When the user selects a MRI study to load, ExactVu reads MR images and Presentation States.
Media Storage Application Profile |
Identifier |
Read/Write |
Compact Disk – Recordable |
||
General Purpose CD-R Interchange |
STD-GEN-CD |
FSR |
General Purpose CD-R Interchange with JPEG |
STD-GEN-CD-JPEG |
FSR |
General Purpose CD-R Interchange with J2K |
STD-GEN-CD-J2K |
FSR |
CT/MR Studies on CD-R |
STD-CTMR-CD |
FSR |
DVD |
||
General Purpose DVD Interchange |
STD-GEN-DVD |
FSR |
General Purpose DVD Interchange with JPEG |
STD-GEN-DVD-JPEG |
FSR |
General Purpose DVD Interchange with J2K |
STD-GEN-DVD-J2K |
FSR |
CT/MR Studies on DVD Media |
STD-CTMR-DVD |
FSR |
USB and Flash Memory |
||
General Purpose USB |
STD-GEN-USB |
FSR |
General Purpose USB with JPEG |
STD-GEN-USB-JPEG |
FSR |
General Purpose USB with J2K |
STD-GEN-USB-J2K |
FSR |
Network Drive |
||
A network drive mapped by a network-shared folder |
ExactVu private |
FSR |
Note: FSR – File-set Reader, FSC – File-set Creator, FSU – File-set Updater
Not applicable.
The following table lists the character sets used in ExactVu.
Character Set Description |
Defined Term |
Character Set |
AE Role |
Default repertoire |
ISO_IR 6 |
ASCII |
Q/R SCU, Storage SCU, FSR |
Latin alphabet No. 1 |
ISO_IR 100 |
ISO 8859/1 |
Q/R SCU, Storage SCU, FSR |
Latin alphabet No. 2 |
ISO_IR 101 |
ISO 8859/2 |
Q/R SCU, Storage SCU, FSR |
Unicode UTF-8 |
ISO_IR 192 |
UTF-8 |
Q/R SCU, Storage SCU, FSR |
ExactVu does not support any specific security measures. It is assumed that ExactVu is used within a secured environment. It is assumed that a secured environment includes at a minimum:
Not applicable.